PICCELL/R (cover removed) is a unique combination of high speed image analysis with powerful wet dispersion
optimised for on-line applications: up to 175 fps with 2048x2048 square pixels, 1 ns exposure time,
displayed: PICCELL/R with flange for the sampling device on top.
Technical specifications and dimensions.
PICCELL/R combines the suspension flow cell of the LIXELL with the high speed image analysis of QICPIC/R in a single robust instrument, specially designed for on-line applications in the process environment. PICCELL/R can directly operate in combination with the representative sampler family TWISTER, sampling probes, the MIXER or user supplied samplers.
Like QICPIC/R in combination with LIXELL it combines innovative new techniques: a family of powerful, high pressure (up to 10 bar) stainless steel (V4A) flow cells with gap widths of 0.2, 0.5, 1, 2 , 4 and 10 mm adapted to the size range; a pulsed light source with an extremely short exposure time of less than 1 ns which enables clear cut images from fastest particles created by the dispersion process (10 m/s would result in only 10 nm of motion blur); a high speed camera suitable to measure up to 175 images per second for high statistical relevance of the measured results and short analysis times. A built-in auto-focus aligns the flow cell along the optical axis for the optimum sharpness of the images.
PICCELL/R covers particle sizes between 1.1 µm and 7.51 mm in five measuring ranges. The primary measurement data is stored in 100,000 primary classes and can be evaluated in individually definable formats. Pre-defined sets of size classes allow an easy adoption to existing measurement specifications.
The WINDOX software offers simultaneous calculations of all relevant size and shape characteristics. A particle gallery in combination with multiple filter functions supports the evaluation of specific results, even if very large amounts of 10 million particles or more are stored.
PICCELL/R fulfils the security requirements of the FDA and is compliant to 21 CFR Rule 11.
|
Particle size and shape analysis from 1.1 µm to 7,510 µm in five measuring ranges. |
|
Direct co-operation with the representative sampler TWISTER, simple sampling probes, the MIXER or user supplied samplers. |
|
Efficient dispersion instead of software-based particle elimination or correction. |
|
Quasi-static image analysis in a high speed particle stream. |
|
Clear outline of particles, independent of the position in the measuring zone. |
|
Definite borders of particles, black outline, transparent particles are no problem. |
|
Very low statistical errors due to very high particle numbers. |
|
User defined class limits out of 100,000 primary size classes, free definition of sampling points, pre-defined sets of class limits are shipped with the software (e.g. HELOS classes, sieve classes, etc.) |
|
A large number of evaluation modes are available, evaluation modes are selectable even after the measurement:
|
|
PICCELL/R fulfils the security requirements of the FDA and is compliant to 21 CFR Rule 11. |
For image analysis applications in the laboratory (off-line) please refer to QICPIC/R.
TWISTER 150 as an example of a representative sampling stage for PICCELL/R
For representative sampling the TWISTER family can be used and directly connected to PICCELL/R.
For homogeneous material simple sampling probes and the MIXER can be used.
Operating at 175 fps, PICCELL/R collects up to 700 Mbytes of data per second. Efficient algorithms evaluate the data transmitted to a PC via two parallel CoaXPress connections (max. 2 x 5 Gbps). In this case, the host computer can be operated over long distances to the measuring location (e. g. 55 m at 3.125 Gbps).
PICCELL/R is fully integrated in and supported by the latest WINDOX software
QICPIC/R with wet disperser LIXELL
For laboratory (i.e. off-line) image analysis applications, please refer to QICPIC/R.